Z-0541-2026: DxFLEX Flow Cytometer Catalog Number C78500 UDI-DI code: 15099590754013 The DxFLEX Flow Cytometer is int

Recall record Z-0541-2026 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0541-2026
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Nov 18, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.

Source facts

Action
On 09/22/2025. an Urgent Medical Device Recall letter was sent via postal mail and/or email informing customers that Beckman Coulter has determined that the DxFLEX software version 2.3.4.37 may crash due to a code error issue.While using the DxFLEX Flow Cytometer instrument for data acquisition using user defined acquisition protocol, the CytExpert for DxFLEX software (version 2.3.4.37) may unexpectedly crash. Due to that, the data acquisition may not be completed according to the preset conditions (i.e., number of events and/or acquisition time). When the software crash is not detected promptly, the sample may be completely consumed. Customers are instructed to: Follow the steps below when running experiments in the instrument until the software is upgraded to the version that will fix the issue: a. Prepare the experiment file (create sample, set location, etc.) b. Switch to first Tube sheet for each task to be acquired c. Save and close the experiment file d. Open the experiment again (DO NOT switch to REPORT sheet after opening experiment) This step will prevent the software from crashing due to this code error issue. e. Perform acquisition f. Analyze data. " Beckman Coulter recommends posting this letter on or near the affected analyzers until the update to the software that addresses this issue is completed. For questions: contact our Customer Support Center " From our website: http://www.beckman.com " By phone: call (800) 369-0333 in the United States.
Initiated
2025-09-22
Root cause
Under Investigation by firm
Distribution
Worldwide - U.S. Nationwide distribution in the states of CA, CT, FL, IL, NY, OH, OR, TN, and TX. The country of India.
Recall number
Z-0541-2026
Recall status
Open, Classified
CDRH record ID
216127
FDA classified
2025-11-18
Recall event number
97624

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗