Test results, low: Sodium, potassium and chloride results on the cobas 6000 ISE unit may exhibit a downward drift in controls and patient samples within the recommended calibration interval of 24 hours. For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.
Source facts
Action
Consignees were notified by letter dated 9/15/08, which provided a work-around for them to use until the problem can be permanently resolved. The letter was entitled "URGENT MEDICAL DEVICE CORRECTION." For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.
Initiated
2008-09-15
Root cause
Package design/selection
Terminated
2009-12-01
Distribution
Nationwide Distribution.
Recall number
Z-0542-2009
Recall status
Terminated
CDRH record ID
73233
FDA classified
2009-01-06
Recall event number
49382