Z-0542-2021: Everest MI XT Tab Removal Tool (i.e. All-In-One Tab Removal Tool)-used to break the tab off the Everest X

Recall record Z-0542-2021 for K2M, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0542-2021
Company in FDA record
K2M, Inc
FDA record creation date
Dec 7, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

All-in-One Tab Removal Tool may cause cracking and/or separating at the threaded interface between tool s metal shaft component and handle, resulting in leakage of Loctite from handle cracks

Source facts

Action
Stryker issued Urgent Medical Deice Letter dated 11/4/20, stating reason for recall, health risk and action to take: 1. Immediately review the Product Recall Notification. Please forward this information to all sales representatives and operations teams within your Branch/Agency. 2. Immediately check your internal inventory to locate the product listed on the attached Business Reply Form and list the quantity of units on the form. Note: Only the Tab Removal Tools with the Catalog number and Lot numbers specified in the Product Affected Table 1 are in scope of this recall. Tab removal tools from other lots should not be returned. 3. Use the Business Reply Form to reconcile any affected product. Complete the Business Reply Form even if there is no affected product identified. 4. Return the enclosed Business Reply Form by email to Spine-RegulatoryActions@Stryker.com or fax to 201.575.4675 to confirm receipt of this notification. 5. Within 5 days of receipt of this notice, mail any affected product to: Stryker Spine 610 Hope Parkway SE Leesburg, VA 20175 a. A red return box will be sent to your specified address. b. Please place the returning part in the red box and return the red box inside of the Fed Ex box provided. Pack only the product listed above. c. Use the FedEx label that has been provided to you by Reverse Logistics. d. Appropriately seal the box. Stryker will contact you to arrange for replacement of your product(s). 6. If you desire additional training associated with these instructions, please contact our Regulatory Compliance team at 201.749.8090 or by email at Spine-RegulatoryActions@Stryker.com.
Initiated
2020-11-04
Root cause
Process change control
Terminated
2024-02-22
Distribution
Nationwide Foreign: Netherlands, Spain, UK, Belgium, Germany, Italy, Norway and Sweden
Recall number
Z-0542-2021
Recall status
Terminated
CDRH record ID
184405
FDA record created
2020-12-07
Recall event number
86733

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗