Z-0542-2024: AirFit F30i Full Face Mask and User Guide

Recall record Z-0542-2024 for ResMed Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0542-2024
Company in FDA record
ResMed Ltd.
FDA classification date
Dec 21, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact

Source facts

Action
On 11/20/23, "Urgent Field Safety Notices" and FAQs were placed on the firms website at www.resmed.com/magnetupdate, and these notices were emailed to healthcare providers, and distributors who were asked to do the following: 1) Provide a copy of this notice and the physician letter to prescribing physicians, and/or any other relevant healthcare professionals, notifying them of the updated labeling (contraindications and warning). 2) Provide a copy of the patient letter to all patients currently using an affected mask with magnets, notifying them of the updated labeling (contraindications and warning). Note, patients are instructed in the patient letter to contact their mask provider for a replacement mask if they are now contraindicated. 3) Provide a replacement mask for contraindicated patients to an alternative mask without magnets in a timely manner. Where an alternative mask is not available, inform patients to consult with their physician. 4) Instruct patients to consult their physician and/or manufacturer of their implant / other medical device if they require additional information on the potential adverse effects of magnetic fields for their particular device (as described in warnings). 5) For alternative mask options, please contact your ResMed or Customer Service representative. 6) Complete and return the customer acknowledgement form via email to magnetresponse@resmed.com For any questions, please contact your local ResMed contact or visit www.resmed.com/contact.
Initiated
2023-11-20
Root cause
Device Design
Distribution
US nationwide distribution including Puerto Rico, Guam, and Virgin Islands. OUS (foreign) countries of: AE, AR, AT, AU, BB, BD, BE, BG, BH, BM, BR, CA, CH, CL, CN, CO, CW, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FO, FR, GB, GF, GG, GL, GP, GR, GT, GY, HK, HN, HR, HU, ID, IE, IL, IM, IN, IR, IS, IT, JE, JO, JP, KE, KG, KR, KW, LB, LK, LT, LV, LY, MA, MC, MM, MQ, MT, MU, MV, MX, MY, NA, NC, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PM, PT, QA, RE, RO, RS, RU, SA, SD, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UY, VE, VN, WF, YT, ZA
Recall number
Z-0542-2024
Recall status
Open, Classified
CDRH record ID
204523
FDA classified
2023-12-21
Recall event number
93499

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗