Z-0542-2026: REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60, IVD in conjunction with a Hon

Recall record Z-0542-2026 for Cellavision AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0542-2026
Company in FDA record
Cellavision AB
FDA classification date
Nov 18, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.

Source facts

Action
On October 10, 2025, CellaVision issued a "Urgent Medical Device Correction" Notification to affected consignees via US Mail and Email. Accompanying the notice was Technical Note TN148. CellaVision asked consignees to take the following action: 1. Identify whether any of your DI-60 analyzer(s) are affected by verifying the serial number(s) against the list of affected units provided in Table 1. 2. If your unit(s) are affected, please implement the following interim measure to mitigate the risk of sample misattribution: a. Before signing each order, confirm that the number of slides in the order matches the number of slides ordered. No order shall contain more slides than prescribed. b. Await contact from Sysmex service personnel, who will arrange installation of the updated CCU firmware. 3. You may continue to use the system(s) in accordance with its intended use, provided the above instructions are followed 4. Please ensure that this Correction Notification is circulated to all users of the device so that they are made aware of the potential issue. 5. Retain this letter until the firmware correction has been implemented. Ensure that the notice is kept in a visible and accessible location. 6. Fill in the customer response form and send via e-mail to whomever you received the Correction Notification from (or to any of the Sysmex America Inc. contacts found on page 1). The response must be sent as soon as possible, and no later than 30 days after you receive this Correction Notification.
Initiated
2025-10-08
Root cause
Software design
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR.
Recall number
Z-0542-2026
Recall status
Open, Classified
CDRH record ID
216699
FDA classified
2025-11-18
Recall event number
97835

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗