Z-0543-2026: MONARCH Bronchoscope. Model Number: MBR-000211-B

Recall record Z-0543-2026 for Auris Health, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0543-2026
Company in FDA record
Auris Health, Inc
FDA classification date
Nov 19, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential that product was leak tested with equipment outside of its expected operating range, resulting in bronchoscopes that may not be susceptible to leaks.

Source facts

Action
Auris Health notified consignees via emailed "URGENT FIELD SAFETY NOTICE - DEVICE RECALL" letter dated 10/17/2025 on about 10/18/2025. Consignees were instructed to cease use of affected units immediately, forward the notification to any applicable personnel as well as customers if further distributed/transferred, and arrange for the return of all affected units. Consignees were also instructed to complete and return the Acknowledgement Form via email to MonarchSupport@its.jnj.com. Please return this form even if you no longer have or use the device subject to this notification. Please return any of the listed bronchoscopes in your possession to Auris Health, Inc. Please contact the MONARCH Support team at +1 (800) 434-0032 or via email at MonarchSupport@its.jnj.com for support in returning any impacted devices. If you have any questions regarding this letter or need support, please contact the MONARCH Support team at +1 (800) 434-0032 or via email at MonarchSupport@its.jnj.com. For Medical Information requests, please visit our website at https://www.jnjmedtech.com/mir.
Initiated
2025-10-18
Root cause
Process control
Distribution
Nationwide distribution to AL, AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KY, MA, MI, MS, MT, NC, NE, NJ, NM, NY, OH, OR, PA, SC, SD, TN, VA, VT, WA, WV.
Recall number
Z-0543-2026
Recall status
Open, Classified
CDRH record ID
216558
FDA classified
2025-11-19
Recall event number
97781

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗