Open original FDA record ↗Recall record
- FDA identifier
- Z-0543-2026
- Company in FDA record
- Auris Health, Inc
- FDA classification date
- Nov 19, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential that product was leak tested with equipment outside of its expected operating range, resulting in bronchoscopes that may not be susceptible to leaks.
Source facts
- Action
- Auris Health notified consignees via emailed "URGENT FIELD SAFETY NOTICE - DEVICE RECALL" letter dated 10/17/2025 on about 10/18/2025. Consignees were instructed to cease use of affected units immediately, forward the notification to any applicable personnel as well as customers if further distributed/transferred, and arrange for the return of all affected units. Consignees were also instructed to complete and return the Acknowledgement Form via email to MonarchSupport@its.jnj.com. Please return this form even if you no longer have or use the device subject to this notification. Please return any of the listed bronchoscopes in your possession to Auris Health, Inc. Please contact the MONARCH Support team at +1 (800) 434-0032 or via email at MonarchSupport@its.jnj.com for support in returning any impacted devices. If you have any questions regarding this letter or need support, please contact the MONARCH Support team at +1 (800) 434-0032 or via email at MonarchSupport@its.jnj.com. For Medical Information requests, please visit our website at https://www.jnjmedtech.com/mir.
- Initiated
- 2025-10-18
- Root cause
- Process control
- Distribution
- Nationwide distribution to AL, AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KY, MA, MI, MS, MT, NC, NE, NJ, NM, NY, OH, OR, PA, SC, SD, TN, VA, VT, WA, WV.
- Recall number
- Z-0543-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216558
- FDA classified
- 2025-11-19
- Recall event number
- 97781