Open original FDA record ↗Recall record
- FDA identifier
- Z-0544-2014
- Company in FDA record
- SpineFrontier, Inc.
- FDA classification date
- Dec 19, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During routine servicing, it was discovered that one 8mmx55mm Reset screw was assembled with an 8mm tulip head and a 9mm screw. The implant should have consisted of an 8mm tulip head and an 8mm screw. A recall was initiated to remove all remaining screws of the same lot from the field to investigate the nonconformity.
Source facts
- Action
- SpineFrontier sent an Voluntary Recall Notification dated November 5,2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. All affected implants are to be returned to SpineFrontier for inspection. SpineFrontier is requiring confirmation that you have received this notice and that you will initiate the immediate return of the implants listed above. IMMEDIATE RESPONSE REQUIRED. Please feel free to contact us at 978-232-3990 Monday through Friday 8:30AM to 5:30PM EST with any questions
- Initiated
- 2013-10-18
- Root cause
- Software design (manufacturing process)
- Terminated
- 2016-08-22
- Distribution
- US Distribution to VA and TX.
- Recall number
- Z-0544-2014
- Recall status
- Terminated
- CDRH record ID
- 123002
- FDA classified
- 2013-12-19
- Recall event number
- 66698