There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
Source facts
Action
On October 18, 2018, Philips notices to their customers via courier service. Customers were advised to take the following actions:
If any unusual noise is heard, discontinue use of the system immediately and contact your Philips representative for support.
Field Order Change (FCO) 72800703 will be deployed to initiate an inspection of the front gantry cover. If the inspection finds evidence of weakening at this connection point, a temporary repair solution will be put in place to secure the cover for continued system use. FCO 72800704 will then be deployed to replace the affected front cover. A Philips Field Service Engineer will contact you to schedule the inspection of your system.
Customers in need of further information or support concerning this issue, please contact your local Philips representative (North America & Canada) 1-800-722-9377.
Initiated
2018-10-18
Root cause
Process control
Terminated
2021-08-03
Distribution
US Nationwide Distribution
Recall number
Z-0544-2019
Recall status
Terminated
CDRH record ID
168978
FDA record created
2018-11-30
Recall event number
81569