Open original FDA record ↗Recall record
- FDA identifier
- Z-0545-2013
- Company in FDA record
- Zimmer, Inc.
- FDA classification date
- Dec 18, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening
Source facts
- Action
- Zimmer sent distributors beginning the week of November 14, 2012, E-mail information requesting all equipment be collected in distributors office before shipment back to Zimmer and replacement. A telephone conference discussing the recall and amended surgical technique was held the week of November 14, 2012, and Urgent Recall notices were sent to Distributors describing the devices; risks; updated surgical technique with directions to hand deliver surgeon/risk manager letters to all affected surgeons and risk managers in hospitals receiving the product. The notices included directions to remove all affected product and ensure risk manager letter is provided. The letter stated a Zimmer Inc. employee will conduct onsite reconciliation and asked return of distribution Spreadsheet to CorporateQuality.Postmarket@zimmer.com within 24 hours of receiving notification. Product is to be returned with certificate of sterilization to Zimmer Product Service Department 1777 West Center St. Warsaw, IN 46580 or Zimmer International Logistics, Eschbach Germany for International Accounts. For questions regarding this recall call 574-372-4807.
- Initiated
- 2012-11-14
- Root cause
- Device Design
- Terminated
- 2014-04-16
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Germany and Europe.
- Recall number
- Z-0545-2013
- Recall status
- Terminated
- CDRH record ID
- 114657
- FDA classified
- 2012-12-18
- Recall event number
- 63690