Open original FDA record ↗Recall record
- FDA identifier
- Z-0545-2014
- Company in FDA record
- ASI Medical, Inc.
- FDA classification date
- Dec 19, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
ASI Medical is initiating a field correction for the SEAC Advanced Dental System because the miniature plastic air tank can potentially rupture.
Source facts
- Action
- ASI Medical sent an Field Correction Action and Safety Notice letter dated November 5, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. REQUIRED FIELD CORRECTION Our field correction will consist of sending you a warning label to be applied to the back service access panel of the cart and an ASME safety pop off valve that will open and exhaust if pressure should reach over 100 psi. This safety pop off valve is designed be easily installed in line between the air compressor and air tank in your system. We are requesting your cooperation in making this field correction.
- Initiated
- 2013-11-05
- Root cause
- Device Design
- Terminated
- 2014-08-11
- Distribution
- Worldwide Distribution - USA (nationwide) including the countries of Austria, Aruba, Saudi Arabia, Canada, Germany, Spain and West Indies.
- Recall number
- Z-0545-2014
- Recall status
- Terminated
- CDRH record ID
- 124413
- FDA classified
- 2013-12-19
- Recall event number
- 66983