Z-0545-2022: CRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detection of Cryptococcal Antig

Recall record Z-0545-2022 for Immuno-Mycologics, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0545-2022
Company in FDA record
Immuno-Mycologics, Inc
FDA record creation date
Jan 31, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive test results may be false positives, which may cause some patients to initiate unnecessary anti-fungal therapy. Samples with negative test results are NOT affected.

Source facts

Action
Customers who received product were contacted via letter Urgent: Medical Device Recall, Cryptococcal Antigen Lateral Flow Assay, dated January 3, 2022, beginning January 3, 2022, to locate and immediately destroy their affected devices. A recall notification and response form was attached to the email that is to be completed by the customers and returned to the firm.
Initiated
2022-01-03
Root cause
Material/Component Contamination
Terminated
2024-05-15
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA.
Recall number
Z-0545-2022
Recall status
Terminated
CDRH record ID
191362
FDA record created
2022-01-31
Recall event number
89370

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗