Z-0545-2024: VITROS 5600 Integrated System Product Code 6802413 and Refurbished Product Code 6802915 Running Software

Recall record Z-0545-2024 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0545-2024
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
Dec 13, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.

Source facts

Action
An URGENT PRODUCT CORRECTION NOTIFICATION letter dated 10/10/23 was sent to customers. REQUIRED ACTIONS " Review the Customer Actions for the Condition Codes described in this notification. " Acknowledge your understanding of this notification by completing the enclosed Confirmation of Receipt form no later than October 18, 2023. " Save this notification with your User Documentation or post this notification by each VITROS 3600/4600/5600/XT 3400/XT 7600 System until this issue is resolved. " If your laboratory has experienced the issue described in this notification and you have not already done so, please report the occurrence to your local Global Services Organization (Previously Ortho Care). Resolution QuidelOrtho is investigating root cause and will communicate again after root cause has been determined. A software update to address this issue is currently in development. Contact Information We apologize for any inconvenience this may cause your laboratory. If you have further questions, please contact our Global Services Organization at 1-800-421-3311.
Initiated
2023-10-10
Root cause
Under Investigation by firm
Distribution
US Nationwide. Global Distribution.
Recall number
Z-0545-2024
Recall status
Open, Classified
CDRH record ID
204196
FDA classified
2023-12-13
Recall event number
93366

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗