Component leads may have intermittent connection to the printed circuit board assembly pads - which could render the device inoperative.
Source facts
Action
On 2/1/06 the firm issued an UREGENT MEDICAL DEVICE CORRECTION letter to customers via certified return receipt. The letter advised customers of the issue and recommended customers inspect their AED at least weekly. The letter also advised customers what to do when readiness indicator displays certain symbols. The firm''s representatives will visit customers to replace the main circuit board assembly.
Initiated
2006-02-01
Root cause
Nonconforming Material/Component
Terminated
2008-03-06
Distribution
Devices were distributed domestically to first responder type consignees and internationally to distributors.
Recall number
Z-0546-06
Recall status
Terminated
CDRH record ID
44027
FDA classified
2006-02-22
Recall event number
34533