Z-0546-2014: Continuous Peripheral Nerve Block Catheter Kit and Set Continuous Peripheral Nerve Block Catheter Kit an

Recall record Z-0546-2014 for Arrow International Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0546-2014
Company in FDA record
Arrow International Inc
FDA classification date
Dec 19, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A labeling inconsistency was discovered in which the lidstock states that the kit contains a stimulator cable with 1.5mm socket, however, the kit contains a stimulator cable with 1.5mm pin.

Source facts

Action
The firm, Arrow International, Inc., sent an "Urgent Medical Device Recall Notification" letter dated October 14, 2013 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use and quarantine any of listed affected products/lots you have in stock; return affected products, customer service representative will contact you with RGA# and instructions for the return of the product, and complete enclosed Recall Acknowledgement form and fax to 1-800-343-2935, Attn: Customer Service, even if you have not affected product. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-800-343-2935.
Initiated
2013-10-09
Root cause
Labeling mix-ups
Terminated
2015-09-21
Distribution
Worldwide distribution: US (nationwide) including states of: MA, LA, CA, GA, KS, TX, PA, NY, MO, WI, NJ, MS, CT, ID, AL, CO, and PR; and Singapore.
Recall number
Z-0546-2014
Recall status
Terminated
CDRH record ID
122660
FDA classified
2013-12-19
Recall event number
66589

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗