Z-0546-2020: Intersomatic JULIET Ti OL cage. For use in orthopedic spinal procedures

Recall record Z-0546-2020 for SPINEART SA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0546-2020
Company in FDA record
SPINEART SA
FDA classification date
Nov 22, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Inner blister pack label is incorrect and does not match the correct external box label.

Source facts

Action
On 02/05/2019, the firm sent an "URGENT: MEDICAL DEVICE RECALL FSCA-04-2019" notification letter via email informing its one U.S. Distributor of the issue of incorrect inner blister pack labeling for Lot Number 4-2899. The customer was instructed to please carefully follow instructions detailed below: 1. Immediately review their inventory and quarantine concerned products if any. 2. For customer who may have further distributed this product to please identify concerned customers and notify them at once of this product recall by using the URGENT: MEDICAL DEVICE RECALL" document. 3. Collect and quarantine all products. 4. Sent back all products with the enclosed Response Form to the Recalling Firm(Chemin du Pr Fleuri 3, 1228 Plan-les-Ouates, Switzerland). E-mail: regulatory@spineart.com. 5. All returned products will be exchanged with batches already available in the Recalling Firm's warehouse. The customers were also informed of the following risks associated with the affected product: The product is always selected before implantation by reading the external label. Consequently, the selection of a wrong size implant can be excluded. At worst the surgeon could be surprised to see that the internal label does not match with the external box label or the product itself regarding the size. A confusion would then be generated. In addition, the patient file information could be partially erroneous as the internal label is usually kept in the records. But the batch number that is good (4-2899) ensures a full traceability.
Initiated
2019-02-05
Root cause
Employee error
Terminated
2021-04-28
Distribution
US: CA OUS: Unknowns
Recall number
Z-0546-2020
Recall status
Terminated
CDRH record ID
177785
FDA classified
2019-11-22
Recall event number
84219

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗