Z-0547-2020: V-Twin, REF 6002-800 or 6002-850, SN 18-3304, IVD, (01)03661540600180

Recall record Z-0547-2020 for Vital Scientific N.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0547-2020
Company in FDA record
Vital Scientific N.V.
FDA record creation date
Nov 22, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During in-house testing it was determined that their is a possibility that during a sample run, the instrument may aspirate a sample from the wrong sample container resulting in a sample result and patient ID mismatch. If not detected, this may lead to an incorrect patient report.

Source facts

Action
On November 15, 2018 the firm sent an Urgent Field safety Notice via E-Mail to all affected consignees. The firm also sent out an updated notifications on November 30, 2018 and January 8, 2019 The customer notification addressed the problem and the conditions in which the problem could happen. the urgent field safety notice asked customers to do follow: 1.ELITechGroup is working with the highest priority on a permanent solution to this problem. Pending this permanent fix, we are requesting your cooperation to carry out this mandatory instruction for each active analyzer used. 2.To guarantee a safe operation of the referenced products the following flowchart and instruction must be followed: A. In case PSID is switched off by default; no further action required B. In case non-barcoded sample container are used, disable the PSID in all instances C. In case barcoded sample containers are used and they are positioned by using the PSID functionality or manually assigned a position, please pay attention to a "b" flag or "barcode not scanned" flag. The b-flag may appear with the result if during the run the barcode could not be verified. Whenever a b-flag is present reject the result and repeat the test for the requested sample. 3. Disabling the PSID: A. Please follow the instructions below to disable the internal barcode reader. Disable PSID when non-barcoded sample containers are used. 4. Disabling the System Barcode A. From the Main Menu, Click F5 Special Functions B. Click F2 Installation C. Select System Parameters D. Set Use System barcode to NO 5.. "b-flag" (barcode not scanned): A. Just before sampling the instrument will scan the barcode. When it cannot read the barcode a b-flag will appear with the result. Whenever the b-flag appears, reject the result and repeat the test. 6. Please check the custom automatic evaluation setting for the b-flag in the [Install menu]. For your convenience you can use the custom automatic evaluati
Initiated
2018-11-15
Root cause
Nonconforming Material/Component
Distribution
NY
Recall number
Z-0547-2020
Recall status
Open, Classified
CDRH record ID
176738
FDA record created
2019-11-22
Recall event number
83910

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗