Skip to content

Public FDA record

Z-0547-2021: Randox Rx Series Mg - Product Usage: For the quantitative in vitro determination of Magnesium in serum, p

Recall record Z-0547-2021 for Randox Laboratories Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0547-2021
Company in FDA record
Randox Laboratories Ltd.
FDA record creation date
Dec 8, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Failing to meet the linearity performance claims quoted on the kit inserts-The linearity is reduced by approximately -30% for Serum/ Plasma and -50% for Urine

Source facts

Action
Randox issued Urgent Medical Device Correction letter via e-mail to US Customers on 13 November 2020 stating reason for recall, health risk and action to take: Discontinue use of and discard any of the above immediately. " Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical. services@randox.com within five working days. Questions or concerns, please contact Randox Technical Services.
Initiated
2020-11-13
Root cause
Under Investigation by firm
Terminated
2021-04-01
Distribution
US Nationwide distribution in the states of GA, NJ.
Recall number
Z-0547-2021
Recall status
Terminated
CDRH record ID
184629
FDA record created
2020-12-08
Recall event number
86827

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗