Open original FDA record ↗Recall record
- FDA identifier
- Z-0547-2022
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA record creation date
- Feb 3, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a calculation error in the measurement when using 2D trace (manual trace) tool. The trace circumference value is overestimated and may potentially result in misdiagnosis of a patient's condition.
Source facts
- Action
- On 11/12/2021, the firm sent a "MEDICAL DEVICE SAFETY CORRECTION" Letter via email and certified US Post to inform customers that within their ultrasound system there is a calibration error in the measurement when using the 2D trace (manual trace) tool. the trace circumference value is overestimated, however, the area value is accurate. Customers are instructed: -To avoid overestimation of circumference measurements, use the default ellipse tool for all circumference measurements. -To avoid overestimation of distance measurements, use default distance tool for all distance measurements. The Recalling Firm will correct this issue with a free-of-charge update to affected ultrasound systems with a new software version. Customer will be contacted by a service engineer to schedule a facility visit to update the system or inform them of a remote update. For questions, send any inquires to: Siemens Healthineers USD Complaints at usd-complaint.team@siemens-healthineers.com
- Initiated
- 2021-11-12
- Root cause
- Software design
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK, AZ, AR, CA, CO, CN, DE, FL, GA, HI, IL, IN, IA, KS, KY, LA, MD, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WV, WI, and WY. The countries of United Arab Emirates, Armenia, Angola, Austria, Australia, Azerbaijan, Bangladesh, Belgium, Bahrain, Brazil, Switzerland, Chile, China, Colombia, Cyprus, Czech Republic, Germany, Denmark, Dominican Rep., Egypt, Spain, Finland, France, United Kingdom, Georgia, Greece, Guatemala, Hong Kong, Croatia, Indonesia, Israel, Iceland, Italy, Japan, Republic Korea, Kuwait, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Poland, Portugal, Qatar, Romania, Serbia, Russian Fed., Saudi Arabia, Sweden, Singapore, Slovakia, Thailand, Turkey, Taiwan, Ukraine, South Africa.
- Recall number
- Z-0547-2022
- Recall status
- Open, Classified
- CDRH record ID
- 190620
- FDA record created
- 2022-02-03
- Recall event number
- 89139