Skip to content

Public FDA record

Z-0548-2024: CereLink ICP Extension Cable (Extension Cable), Model No. 826845 Used with Codman CereLink Intracranial

Recall record Z-0548-2024 for Integra LifeSciences Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0548-2024
Company in FDA record
Integra LifeSciences Corp.
FDA classification date
Dec 14, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Firm is expanding the existing recall RES 90457 for the Cerelink, initiated in June 2022, to include all CereLink ICP Extension Cables. The updated corrective action includes a design change to the extension cable to address the "out of range readings" issue.

Source facts

Action
On October 5, 2023, the firm issued an updated letter, intended to replace the notification sent on June 22, 2022 for the out-of-range readings issue (RES 90457). As an output of the root cause and failure investigation, the design of the extension cable is being modified to incorporate a solution to address the "out of range readings" issue. The firm instructed customers to return all existing CereLink ICP Extension Cables in their inventory. This includes all CereLink ICP Extension Cables included with the monitor AND all CereLink ICP Extension Cables provided separately. Customers should check their inventory and quarantine any affected product. Once the firm receives the customer's response form, the firm will provide instructions to return the product. Customers with questions about this recall should contact their Account Manager.
Initiated
2023-10-05
Root cause
Device Design
Distribution
US distribution nationwide to the following states: AL AR AZ CA CO CT DC DE FL GA IA KY LA MA ME MN ND NE NH NJ NY OK PA RI TN TX UT VA WI WV OUS distribution to Australia/New Zealand, Canada, and Europe/Middle East/Africa.
Recall number
Z-0548-2024
Recall status
Open, Classified
CDRH record ID
204305
FDA classified
2023-12-14
Recall event number
93084

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗