Open original FDA record ↗Recall record
- FDA identifier
- Z-0549-2014
- Company in FDA record
- Qualigen Inc
- FDA classification date
- Dec 19, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Qualigen initiated this recall because the FastPack kit may cause lower than expected results due to a manufacturing defect.
Source facts
- Action
- Qualigen, Inc., notified customers of the recall by calling customers via telephone (with a telephone script) on 11/08/2013. The telephone (script) included the reason for recall, product description with codes, instructions and contact information. The customers were instructed to examine your inventory and discontinue use of the kits immediately. Qualigen will send you replacement kits and provide you with a shipping label to return all of the defective kits including any that were partially used. Feel free to contact System Support at 760.579.6900 if you have any questions or require assistance with returning your kits.
- Initiated
- 2013-11-08
- Root cause
- Process change control
- Terminated
- 2014-09-18
- Distribution
- US Distribution to states of: AL, CA, KS, KY, IN, MA, NC, NY, NV, OK, SC, TN, TX, UT, VA, and WA.
- Recall number
- Z-0549-2014
- Recall status
- Terminated
- CDRH record ID
- 123729
- FDA classified
- 2013-12-19
- Recall event number
- 66842