Z-0549-2021: Canon Aquilion Prime SP, Multislice Helical CT Scanner, TSX-303B/8L - Product Usage: intended to acquire

Recall record Z-0549-2021 for Canon Medical System, USA, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0549-2021
Company in FDA record
Canon Medical System, USA, INC.
FDA record creation date
Dec 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing resulting in rescanning and reinjection of contrast medium.

Source facts

Action
On 11/9/2020, Canon issued a "Urgent: Medical Device Correction" letter to all affected customers via facsimile. In addition, to informing consignees about the recall, the firm ask affected consignees to take the following actions: 1. Until corrective action has been taken, in order to avoid the problem, please use the system with setting InstaView OFF according to the last notification (dated June 2020) Subject: Abnormality in Realtime Reconstruction Images. Please see Attachment A. 2. Please share this information with all users and reviewing radiologists as well as the clinical engineering or biomedical group at your facility. 3. If you have any questions, please feel free to contact Terry Schultz Regulatory Affairs Manager at (800) 421-1968 or your local Canon service representative at (800) 521-1968.
Initiated
2020-11-09
Root cause
Software design
Terminated
2022-12-16
Distribution
US Nationwide distribution. OUS: Unknown
Recall number
Z-0549-2021
Recall status
Terminated
CDRH record ID
184721
FDA record created
2020-12-09
Recall event number
86864

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗