Z-0549-2024: Ivenix Infusion System (IIS), Large Volume Pump LVP-0004

Recall record Z-0549-2024 for Fresenius Kabi USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0549-2024
Company in FDA record
Fresenius Kabi USA, LLC
FDA classification date
Dec 20, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The device may experience mechanical interference on the Fluid Valve pins, which will trigger a Pump Problem alarm. The issue may lead to delay or interruption of therapy. Depending on the therapy and duration of delay/interruption, the issue may lead to serious harm or death.

Source facts

Action
On November 29, 2023, the firm began notifying affected customers via" URGENT-Medical Device Recall" letter dated November 28, 2023. Customers were provided with precautions to take until the affected LVPs are able to be returned for repair or replacement of internal hardware. Customers will be contacted by a Fresenius Kabi service representative to create a plan for addressing any affected LVPs at their facility. This effort is expected to begin in January 2024. Fresenius Kabi will repair LVPs that may encounter mechanical interference causing the Fluid Valve Actuation System not to move properly. If the LVP cannot be repaired, it will be replaced at no cost to the customer. If you have any questions contact Customer support at, email: Ivenix_support@fresenius-kabi.com or phone at 1-978-775-8100.
Initiated
2023-11-29
Root cause
Process change control
Distribution
US distribution to states of: CA, CO, NJ, UT, and WI.
Recall number
Z-0549-2024
Recall status
Open, Classified
CDRH record ID
204472
FDA classified
2023-12-20
Recall event number
93474

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗