Z-0550-2025: GE Proteus XR/A radiographic system

Recall record Z-0550-2025 for GE Healthcare (China) Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0550-2025
Company in FDA record
GE Healthcare (China) Co., Ltd.
FDA classification date
Nov 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Steel cables in the Wall Stand for certain Proteus XR/a systems serviced by GE HealthCare have not been replaced at the intervals detailed in the Preventative Maintenance Action List.

Source facts

Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 10/24/24 was sent to customers. Actions to be taken by Customer/User " You can continue to use the system for table exams without limitations. " You can continue to use the wall stand for exams where the patient does not have any part of their body underneath the bucky. " Do not perform exams where the patient has any body part underneath the bucky until the Wall Stand steel cable has been inspected and if necessary, has been replaced. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to RECALL.17148@gehealthcare.com GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2024-10-24
Root cause
Equipment maintenance
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-0550-2025
Recall status
Open, Classified
CDRH record ID
210925
FDA classified
2024-11-25
Recall event number
95674

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗