Z-0551-06: Baxter I-Pump Pain Management System infusion pump, product codes 2L3107, 2L3107K and 2L3107R, Baxter Hea

Recall record Z-0551-06 for Baxter Healthcare Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0551-06
Company in FDA record
Baxter Healthcare Corp.
FDA classification date
Apr 26, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Delivery of unrequested PCA doses of medication due to damaged PCA cords, partial sticking of PCA button, and fluid ingress into the pump or PCA button.

Source facts

Action
Baxter sent Urgent Device Correction letters dated 12/21/05 to the direct accounts on the same date via first class mail. The letters informed the accounts of the potential for unrequested PCA doses due to an electrical short in the PCA circuit simulating repeated pressing of the PCA button. The unrequested bolus doses will not exceed the programmed prescription limits. The letter listed three conditions identified that can produce an electrical short in the PCA circuit and simulate repeated pressing of the PCA button: * Damaged PCA cord or button, * Partial sticking of the PCA button, and * Fluid ingress into the pump or PCA button. Information was provided to reduce the potential for these conditions from occurring during use of the pumps. Customers were asked to ensure that prior to or after each use the PCA cord is intact and has no cuts or missing insulation, and that the pump connector and the button are securely attached to the cord. They were also asked to immediately take the pump and PCA cord out of service for repair if the pump displays 'Release the PCA Button' and the PCA button is not being intentionally pressed, as there may be a mechanical or electronic fault in the PCA button. They were also asked two avoid getting liquids inside the pump as permanent damage may result. Baxter recommended that the institutions implement a regular preventative maintenance program and periodic replacement of the PCA cords based on their own specific usage patterns. Any questions were directed to the Center for One Baxter at 1-800-422-9837. Baxter sent a follow-up recall letter to the accounts on 6/21/06, providing them with an Operator's Manual Addendum for the AP II Infusion Pump and the PCA II Infusion Pump, with the warnings to ensure that prior to or after each use that the PCA cord is intact and has no cuts or missing insulation, and that the pump connector and the button are securely attached to the cord; to avoid getting liquids inside the pump; and
Initiated
2005-12-21
Root cause
Other
Terminated
2008-11-03
Distribution
Nationwide including Puerto Rico, and internationally to Canada, Chile, China, Hong Kong, Korea, Kuwait, Malaysia, Taiwan and Saudi Arabia
Recall number
Z-0551-06
Recall status
Terminated
CDRH record ID
43325
FDA classified
2006-04-26
Recall event number
34244

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗