Sample/Patient Mis-identification: Two issues identified related to the ACCELERATOR Decision Manager (ADM), to include: 1) Sample Identification (SID) and/or Patient Identification (PID) numbers that contain more than 12 characters are truncated to the final 12 characters by the ADM; 2) for SIDs using alpha characters that are case sensitive, SIDs with lower case characters are transmitted by the
Source facts
Action
Firm notified consignees via Product Correction, Immediate Action Required letter(s) on 10/25/07. Consignees were notified of software issues and informed they would be contacted by an Abbott representative to schedule installation of software upgrade.
Initiated
2007-10-25
Root cause
Software design
Terminated
2008-04-01
Distribution
Nationwide:Distributed to hospitals/laboratories in IL, PA, TN, TX, UT, VA and WV.
Recall number
Z-0551-2008
Recall status
Terminated
CDRH record ID
66375
FDA classified
2008-01-25
Recall event number
45873