Z-0551-2022: Spiration Valve System (SVS) Valve in Cartridge -A one-way endobronchial valves indicated for adult patie

Recall record Z-0551-2022 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0551-2022
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
Feb 1, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Mislabeled Patient Charge Label indicating model number SVS-V7-00 (7mm) instead of SVS-V9-00 (9mm)

Source facts

Action
Gyrus ACMI Inc. (Olympus) issued Urgent Medical Device Correction Letter dated 12/23/21. Letter states reason for recall, health risk and action to take: Inspect your inventory for the referenced device and identify any devices with the model number and lot number specified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. Please contact Customer Service 1-888-524-7266, option 1, to obtain a Return Material Authorization. Olympus will issue a credit or replacement to your facility for your affected product. 2. Review your facility s use of the patient charge labels. If your facility has used these mislabeled patient charge labels in the patient s chart, or scanned into electronic medical records and/or patient implant cards please request replacement labels by indicating the quantity in the Recall Portal (see Step 3). Olympus will ship replacement labels for you to use in correcting your patient records. In the case of labels sent with patients at discharge for their personal records, include an explanation of the situation as described above. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number "0403 and provide your contact information as indicated in the portal. As you are most likely aware and as provided in the labeling for the product, for any subsequent procedures with that patient, the SVS endobronchial valve size selection is based on patient anatomy using a sizing kit (SVS-VSK-06) and the fit is verified by a trained clinician after placement according to the product s instruction manual. As a result, the mislabeled patient charge label was determined to not adversely affect medical treatment. However, Olympus recognizes that there may be confusion about what product size was previously used and is therefore sending this notice to you. Olympus requests you to report any complaints, including product l
Initiated
2021-12-23
Root cause
Labeling Change Control
Distribution
US Nationwide distribution in the states of AZ, CA, MA, NJ, PA, SC, TX, WI.
Recall number
Z-0551-2022
Recall status
Open, Classified
CDRH record ID
191288
FDA record created
2022-02-01
Recall event number
89333

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗