Z-0551-2024: Change Healthcare Stratus Imaging PACS 1.4 and Stratus Imaging Share 1.4, a component of Stratus Imaging

Recall record Z-0551-2024 for CHANGE HEALTHCARE CANADA COMPANY. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0551-2024
Company in FDA record
CHANGE HEALTHCARE CANADA COMPANY
FDA classification date
Dec 14, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Under certain circumstances, information from HL7 messages received by Stratus PACS/Imaging Share systems has been incorrectly associated with another patient's study.

Source facts

Action
The recalling firm issued an Urgent Field Safety Notice dated 11/1/2023 which explained the issue that under certain circumstances, one or more of the following three issues may occur: (1) Existing radiology reports may be associated with the wrong patient or study; (2) radiology reports may be missing from imaging studies; and (3) patient metadata in DICOM images may be overwritten with metadata from another patient or study. The notice reports the issue has only existed in the field since it's implementation on 10/26/2023 and the recalling firm is in the process of identifying a list of impacted patient records and the consignee will be provided an update within 24 hours. The circumstances under which the issue occurs are still being investigated. Risk to patient is provided along with the immediate actions to be taken to reduce the risk to patient. The actions include if a radiology report is expected and not available in systems downstream from Imaging Share, customers are advised to contact the radiology department where the reporting originally occurred. Customers are advised to review the audit log for instances of "Exam info modified" to determine if changes to study details have been made in error. In situations where patient demographics have been changed, this audit log will provide the original values for impacted patient records. As this issue extends beyond Radiology department workflows, customers are advised to have heightened awareness to correlate metadata with other systems to ensure accuracy of records. As Imaging Share can distribute images and reports to downstream systems, customers are also advised to review any potentially impacted patient records as identified in the audit log in all downstream systems and workflows. The notice informs the customer that the recalling firm has implemented a fix to prevent further impact to patient records in version 1.4.0 hotfix 02. Regarding recommended actions for the customer to take on potent
Initiated
2023-11-01
Root cause
Software design
Distribution
US Nationwide distribution. There was no military/government distribution.
Recall number
Z-0551-2024
Recall status
Open, Classified
CDRH record ID
204363
FDA classified
2023-12-14
Recall event number
93451

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗