Z-0552-2010: Philips Avalon Fetal Monitor FM 50 Catalog Number: M2705A.

Recall record Z-0552-2010 for Philips Healthcare Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0552-2010
Company in FDA record
Philips Healthcare Inc.
FDA classification date
Dec 17, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Potential for inaccurate ultra-sound derived fetal heart rate readings in Philips Avalon Fetal Monitors. Although the types of inaccuracies described in these complaints reflect known limitations inherent to ultrasound fetal heart rate monitoring, the frequency of such complaints is greater for these devices than for the previous generation Philips Series 50 fetal monitors. Clinical decisions bas

Source facts

Action
Philips notified customers with an Urgent Medical Device Recall notice and Instructions for Use Addendum on 11/23/09 via UPS. Philips is providing the Addendum to reinforce the Avalon Instructions for Use regarding: " Recognizing these phenomena as they relate to the Avalon Fetal monitor and fetal monitors in general; and " Understanding the differences between the Philips Avalon Fetal Monitors and their predecessors, the Philips Series 50 Fetal Monitors, as they relate to the display of ultrasound-derived FHR. Philips is providing access to application specialists who are on call 24 hours a day to assist users with questions related to device use by calling 1-800-722-9377, press <3> then <1> for patient monitoring application issues and ask to speak with an application specialist. Customers will be requested to incorporate the information contained in the Urgent Medical Device Recall notice and the Addendum to the Instructions for Use with the existing Avalon labeling (Instructions for Use) provided with the product. Customers will be asked to complete an enclosed Confirmation Form indicating that they have received and disseminated this information, and then return the completed form to Philips. Questions about this field action or reports of product problems should be directed to the Philips Response Center at 1-800-722-9377.
Initiated
2009-11-23
Root cause
Process design
Terminated
2020-07-30
Distribution
Nationwide (US and Puerto Rico). Foreign: Afghanistan Algeria Argentina Australia Austria Bahrain Belgium Bolivia Bosnia and Herzegovina Botswana Brazil Bulgaria Canada Chile China Colombia Costa Rica Croatia Czech Republic Denmark Dominican Republic Ecuador Egypt El Salvador Estonia Finland France French Guiana Gabon Germany Greece Guadeloupe Guatemala Hong Kong Hungary Iceland India Indonesia Ireland Israel Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Latvia Lebanon Libyan Arab Jamahiriya Lithuania Luxembourg Macau Malaysia Martinique Mauritius Mexico Moldova Morocco Namibia Netherlands Netherlands Antilles New Caledonia New Zealand Norway Oman Pakistan Panama Peru Philippines Poland Portugal Puerto Rico Qatar Reunion Romania Russia Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sweden Switzerland Syrian Arab Republic Taiwan Thailand Tunisia Turkey Turks and Caicos Islands Ukraine United Arab Emirates United Kingdom Uzbekistan Venezuela Viet Nam Zambia
Recall number
Z-0552-2010
Recall status
Terminated
CDRH record ID
86827
FDA classified
2009-12-17
Recall event number
53889

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗