Z-0552-2016: Airways Development LLC WaterPAP Positive Airway Pressure Device Reference #WP-7700.

Recall record Z-0552-2016 for Airways Development LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0552-2016
Company in FDA record
Airways Development LLC
FDA classification date
Dec 24, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Airways Development LLC has received a complaint about a canister leaking from WaterPAP Lot 8605A. During a review with their supplier it could not be determined if this was an isolated incident or if the whole lot is impacted.

Source facts

Action
Airways Development Inc. sent Product Alert notices dated 9/24/2015 to their customers via email. The alert informed customers about the related product issue and how to identify affected product. Airways Development is replacing any affected product left in their customers stock. Customers were also instructed to notify their customers if the product was further distributed. A point of contact was provided if the customers had any related questions.
Initiated
2015-09-24
Root cause
Nonconforming Material/Component
Terminated
2016-05-09
Distribution
Nationwide
Recall number
Z-0552-2016
Recall status
Terminated
CDRH record ID
142352
FDA classified
2015-12-24
Recall event number
72848

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗