GE HealthCare has become aware of the potential that incorrect torque has been applied to certain bolts of the Definium Tempo and Definium Tempo Pro products which may result in a potential fall of elements of the Overhead Tube Suspension (OTS) assembly.
Source facts
Action
GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION to its consignees on 11/07/2023 via FedEx. The notice explained the issue, hazard, and recommended the use of alternative device for patient exams until a service appointment can be scheduled to address the issue.
Questions or concerns: Contact 1-800-467-1171 or local Service Representative
Initiated
2023-11-07
Root cause
Process design
Distribution
US Nationwide distribution.
Recall number
Z-0552-2024
Recall status
Open, Classified
CDRH record ID
204475
FDA classified
2023-12-14
Recall event number
93479
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.