Skip to content

Public FDA record

Z-0552-2025: Merlin@home with the MerlinOnDemand capability Transmitter REF EX1150 Merlin@home with the MerlinOnDeman

Recall record Z-0552-2025 for Abbott Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0552-2025
Company in FDA record
Abbott Medical
FDA classification date
Nov 26, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Due to heart transmitter not being able to upgrade to the current software version in the field after experiencing a power interruption during a prior over-the-wire software download.

Source facts

Action
On 10/25/2024, the firm initiated notifying customers in person/verbally regarding a limited number of Merlin@home Transmitters with MerlinOnDemand Capability that are unable to upgrade to the latest version of software. This issue was caused by a power interruption during a past software updated attempt. Abbott will be retrieving and replacing the affected Transmitters. The replacement Transmitter will have the latest software update version 9.0rev1 For questions or further assistance, speak with or contact the Abbott representative that is on-site replacing the Transmitter, or contact the Public Contact at 408-845-3427
Initiated
2024-10-01
Root cause
Software design
Distribution
Worldwide - U.S. Nationwide distribution in the states of ALABAMA, ALASKA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, ILLINOIS, INDIANA, KENTUCKY, MICHIGAN, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEW JERSEY, NEW YORK, NORTH CAROLINA, OHIO, OREGON, PENNSYLVANIA, SOUTH CAROLINA, TENNESSEE, TEXAS, VIRGINIA, WASHINGTON and WISCONSIN. The country of Australia.
Recall number
Z-0552-2025
Recall status
Open, Classified
CDRH record ID
210583
FDA classified
2024-11-26
Recall event number
95557

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗