Fluid Leaks-The T fitting located in the Symphony Clear system of the Symphony High Volume Slide Stainer needs to be larger to securely hold the tubing in place to eliminate any possible fluid leak due to the tubing movement.
Source facts
Action
The end user customers were verbally contacted starting in November 2007 to set up a scheduled visit to replace the fitting.
Initiated
2007-10-30
Root cause
Component design/selection
Terminated
2008-04-14
Distribution
Worldwide-USA, France and Japan
Recall number
Z-0553-2008
Recall status
Terminated
CDRH record ID
66420
FDA classified
2008-01-25
Recall event number
45891