The low and high controls for the product are returning results above the assigned values.
Source facts
Action
Binding Site sent an Urgent Field Safety Notice dated December 22, 2017.
Advise on action to be taken by the User:
" The new QC values detailed below should be used for the remainder of the shelf life of this Rheumatoid Factor Kit for use on SPAPLUS.
"Input the revised QC values into your SPAPLUS instrument
"Sign and return the TSWS18 E-Back Form accompanying this notification within 5 business days
"Refer to Section 9 of the Product Insert SIN282(.A) if QCs are out of range following reassignment. For further questions, please call (858) 453-9177.
Initiated
2018-01-04
Root cause
Under Investigation by firm
Terminated
2020-06-09
Distribution
US Distribution was made to CA. There was no foreign/military/government distribution.
Recall number
Z-0553-2018
Recall status
Terminated
CDRH record ID
160910
FDA record created
2018-02-08
Recall event number
78934