Skip to content

Public FDA record

Z-0553-2018: Rheumatoid Factor (RF) Kit for use on SPAPLUS, Product #LK151.S.A Intended for the quantitative in-vitro

Recall record Z-0553-2018 for The Binding Site Group, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0553-2018
Company in FDA record
The Binding Site Group, Ltd.
FDA record creation date
Feb 8, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The low and high controls for the product are returning results above the assigned values.

Source facts

Action
Binding Site sent an Urgent Field Safety Notice dated December 22, 2017. Advise on action to be taken by the User: " The new QC values detailed below should be used for the remainder of the shelf life of this Rheumatoid Factor Kit for use on SPAPLUS. "Input the revised QC values into your SPAPLUS instrument "Sign and return the TSWS18 E-Back Form accompanying this notification within 5 business days "Refer to Section 9 of the Product Insert SIN282(.A) if QCs are out of range following reassignment. For further questions, please call (858) 453-9177.
Initiated
2018-01-04
Root cause
Under Investigation by firm
Terminated
2020-06-09
Distribution
US Distribution was made to CA. There was no foreign/military/government distribution.
Recall number
Z-0553-2018
Recall status
Terminated
CDRH record ID
160910
FDA record created
2018-02-08
Recall event number
78934

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗