Potential metal burrs in the holes of the device preventing
the appropriate drill and transverse pins from being inserted
Source facts
Action
Zimmer Biomet initiated recall by letter on 10/24/18 to distributors and hospitals advising of the problem and action to be taken-locate quarantine and return product. Complete Inventory Return Certification Form. Questions or concerns call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday.
Initiated
2018-10-24
Root cause
Employee error
Terminated
2020-04-14
Distribution
US Nationwide Distribution in the states of IN, DC, MD
Recall number
Z-0553-2019
Recall status
Terminated
CDRH record ID
168863
FDA record created
2018-12-04
Recall event number
81513