Z-0553-2021: ELI 380 Electrocardiograph - Product Usage: intended to be a high-performance, multichannel resting elect

Recall record Z-0553-2021 for WELCH ALLYN, INC/MORTARA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0553-2021
Company in FDA record
WELCH ALLYN, INC/MORTARA
FDA record creation date
Dec 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The radio within the device can become disassociated with the wireless access point.

Source facts

Action
The products associated with this Urgent Medical Device Correction letter were manufactured between June 03, 2019 and July 27, 2020. A list of the affected part numbers is provided in Table 1. Action to be taken by the user: o If an ECG test is needed when there is no wireless connection, follow the user manual and acquire an ECG using the STAT ECG mode. The operating software installed on the ELI380 Resting Electrocardiograph requires to be upgraded to version 2.4.3 in order to address this issue. Included with this Urgent Medical Device Correction letter is a USB flash drive with the update software to be used to update your affected units identified in Appendix C. o Enclosed in Appendix C is a list of impacted devices and serial numbers. Locate the devices in your possession and upgrade the software as soon as possible following the instructions in Appendix B or using the instructions provided on the USB flash drive. After the software is successfully upgraded, complete the enclosed Urgent Medical Device Correction Confirmation Form in Appendix C. Sign and return the form to mod1319@hillrom.com. Note: Instructions are provided in Appendix A to locate the serial number on the device if needed. o If technical assistance is needed during the installation process, please contact the appropriate regional technical support via phone or email using the contact information provided in the Contact Reference Person table below. o This notice should be sent to all members of your organization who need to be aware and/or to any organization where the potentially affected devices have been transferred. Transmission of this Urgent Medical Device Correction Letter: Please transfer this notice to other organization as appropriate. Please maintain awareness of this notice for an appropriate period to ensure effectiveness. Please note that this Urgent Medical Device Correction Notice is being made with the knowledge of the Food and Drug Administration (FDA). Adverse re
Initiated
2020-11-11
Root cause
Software design
Terminated
2023-08-08
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IN, KS, KY, MA, MD, MI, MO, MT, NC, ND, NH, NM, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, WA, WI, WV, WY and the countries of Australia, Canada, Czech, Finland, GB, Germany, Italy, Kuwait, Netherlands, New Zealand, Norway, Qatar, Switzerland, United Arab Emirates, United Kingdom.
Recall number
Z-0553-2021
Recall status
Terminated
CDRH record ID
184382
FDA record created
2020-12-09
Recall event number
86716

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗