Skip to content

Public FDA record

Z-0553-2022: ARIES HSV 1&2 Assay, Catalog/REF# 50-10017

Recall record Z-0553-2022 for Luminex Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0553-2022
Company in FDA record
Luminex Corporation
FDA record creation date
Feb 2, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.

Source facts

Action
On 11/17/2021, the firm notified affected customers via letter titled, "URGENT: MEDICAL DEVICE RECALL ARIES HSV 1&2 Assay." The letter indicated, if you observe any leaks, please contact Luminex Global Support Services at support@luminexcorp.com. Please inform all those who need to be aware within your organization of this potential product performance issue. Please contact Luminex Global Support Services at support@luminexcorp.com at your earliest convenience if you have a Kit with an affected lot to arrange for its replacement, and to confirm the disposal of any remaining Kits in your inventory. We ask that you complete and return the attached acknowledgement letter so that we can confirm the destruction of recalled lots. Luminex Global Support Services, 1-877-785-2323 (U.S. and Canada}, +1-512-381-4397 (Outside U.S. and Canada), support@luminexcorp.com
Initiated
2021-11-17
Root cause
Nonconforming Material/Component
Terminated
2024-07-08
Distribution
Worldwide distribution - US Nationwide distribution in the states of New Jersey, California, Connecticut, Minnesota, Missouri, New Jersey, New Mexico, Pennsylvania, Tennessee, Washington, and Wisconsin) and the country of Malaysia.
Recall number
Z-0553-2022
Recall status
Terminated
CDRH record ID
190927
FDA record created
2022-02-02
Recall event number
89244

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗