A variation in suppliers welding processes, at the inner section of the hydraulic lift column, of the surgical table, may result in the table top becoming unstable, which could result in injury to patients or staff.
Source facts
Action
A certified letter was sent to all affected consignees on February 6, 2007 notifying them of the issue and informing them of the recall and the replacement of the affected tables with new tables.
Initiated
2007-02-06
Root cause
Other
Terminated
2009-09-11
Distribution
USA and Canada
Recall number
Z-0554-2007
Recall status
Terminated
CDRH record ID
50396
FDA classified
2007-02-23
Recall event number
37349