Open original FDA record ↗Recall record
- FDA identifier
- Z-0554-2008
- Company in FDA record
- Varian Medical Systems Inc
- FDA classification date
- Jan 29, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Erroneously Resetting: RPM System version 1.7.3, software failure can occur that affects the gated radiotherapy treatment when phase-based gating is used. RPM 1.7 erroneously resets the phase-based gating thresholds to system default during the treatment session, if the user accesses the Session options during the treatment session. This can lead to difference between the planned and delivered
Source facts
- Action
- Varian sent an Urgent Medical Device Correction letter on September 26, 2007, to all its direct consignees, informing them that the RPM version 1.7 may reset the phase thresholds if Sessions options are accessed during treatment. A Service Representative will be scheduling a visit to user site to perform the modification requested to address the problem, when a corrected software version is available.
- Initiated
- 2007-09-26
- Root cause
- Software design
- Terminated
- 2009-09-30
- Distribution
- Worldwide including countries of: Ireland, Italy, Spain, UK, Denmark, Finland, France, and Canada
- Recall number
- Z-0554-2008
- Recall status
- Terminated
- CDRH record ID
- 65722
- FDA classified
- 2008-01-29
- Recall event number
- 45519