A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate.
Source facts
Action
Certified letters were sent to all affected customers on 07/14/2017, to offer a replacement of the battery charger.
Initiated
2017-07-14
Root cause
Component design/selection
Terminated
2018-12-17
Distribution
Distributed in 26 states: AL, CA, CO, FL, GA, IA, IN, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, PA, SD, TX, UT, WA, and WY; and Puerto Rico.
Recall number
Z-0554-2018
Recall status
Terminated
CDRH record ID
161211
FDA record created
2018-02-08
Recall event number
79005