Skip to content

Public FDA record

Z-0554-2025: Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Recall record Z-0554-2025 for Shimadzu Corp. Analytical & Measuring Instruments Div.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0554-2025
Company in FDA record
Shimadzu Corp. Analytical & Measuring Instruments Div.
FDA classification date
Nov 26, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.

Source facts

Action
On 10/9/24, field safety notices were mailed to customers asking them to do the following: Check the Serial Numbers, and if the instruments are affected by this issue, contact your regional office. The corrective actions to be taken to affective products are (1) Add a cover part to the main board. (2) Update the SIL-40 firmware to Ver. 1.16 or above. For questions, please call 800-477-1227 between the hours of 8:30-5:00 EST or visit www.ssi.shimadzu.com.
Initiated
2024-10-09
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.
Recall number
Z-0554-2025
Recall status
Open, Classified
CDRH record ID
210844
FDA classified
2024-11-26
Recall event number
95633

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗