Skip to content

Public FDA record

Z-0556-05: The Triathlon Posteriorly Stabilized Femoral Component.

Recall record Z-0556-05 for Stryker Howmedica Osteonics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0556-05
Company in FDA record
Stryker Howmedica Osteonics Corp.
FDA classification date
Feb 24, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The fatigue testing of the Triathlon PS Femoral Component does not consistently meet the requirements of the finite element analysis predicted load as delineated in the 510(k) submittal.

Source facts

Action
Notification letters and Product Accountability Froms have been sent via Fedex on 10/27/2004 with return receipt.
Initiated
2004-10-27
Root cause
Other
Terminated
2005-05-31
Distribution
The product is distributed to its branches/agencies and to hospitals nationwide.
Recall number
Z-0556-05
Recall status
Terminated
CDRH record ID
35769
FDA classified
2005-02-24
Recall event number
30379

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗