Skip to content

Public FDA record

Z-0556-2007: OEC 9800 Fluoroscopic X-ray System, Model Number A349855, GE Healthcare, Surgery.

Recall record Z-0556-2007 for GE OEC Medical Systems, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0556-2007
Company in FDA record
GE OEC Medical Systems, Inc
FDA classification date
Mar 22, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

During routine service, a cover may have been installed without the required proper lead shielding.

Source facts

Action
Consignees were notified by telephone on 02/09/2007 and told to discontinue use of the units until they could be inspected by GE personnel. GE personnel visited all 8 sites on 02/12/2007.
Initiated
2007-02-09
Root cause
Other
Terminated
2007-06-26
Distribution
Nationwide to AL, AZ, IA, OH, TX, VA, and WV.
Recall number
Z-0556-2007
Recall status
Terminated
CDRH record ID
50344
FDA classified
2007-03-22
Recall event number
37322

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗