Z-0556-2021: MH-200S, system with Ceiling Suspended C-arm Support - Product Usage: intended to be used for cardiac an

Recall record Z-0556-2021 for Shimadzu Medical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0556-2021
Company in FDA record
Shimadzu Medical Systems
FDA record creation date
Dec 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has identified a problem with the control software for the celling arm of the X-Ray System . This software issue could cause the ceiling arm of the device to move after the operator has released the movement lever. This could result in possible collision or harm to the patient or system operator.

Source facts

Action
On 11/18/2020 Shimadzu sent an "Urgent Medical Device Recall Notice" to all affected consignees. In addition to informing consignees about the recall the firm provided the following information. 1. Shimadzu has released a mandatory Service Bulletin on applicable products for the update of the positioner software. The update of the software and the testing of the positioner movement should be performed by your Authorized Shimadzu Service provider. 2. Your authorized Distributor has acknowledged receipt the Service Bulletin (XR-202026) and the recall Customer Letter. They have been instructed to provide you a copy of this letter so that you may understand the scope of the recall. Necessary software for this recall and instruction to complete have been sent to your Distributer and they should be scheduling time to perform the corrective actions. 3. Once the corrective work has been completed your Distributor will report to Shimadzu that the work has been completed so that we may track and inform the appropriate Regulatory Agencies of its completion. Actions to be taken by customer: A. Please check the list of systems affected by this recall. If your site is on the applicable list, please pay attention to the following points until the corrective work has been completed. . (1) This event does not occur in any position other than the Left or Right diagonal position, so please avoid using this product in diagonal positions. (2) When using the product in the Left or Right diagonal position, please use the direct memory function and encourage C-arm operation without using the lever. (3) If the C-arm starts to operate unintentionally, press the red "Stop Switch" button immediately. B. If you feel there is a problem with the c-arm movement, please discontinue use of the system and contact your local service provider. C. Please complete the Medical device Recall Return Response form and submit to drodriguez@shimadzu-usa.com D. Shimadzu
Initiated
2020-11-18
Root cause
Software design
Distribution
US Nationwide distribution including in the states of KY, CA, MI, OH, CT, MI, NY, SC, NC,IN, IL, TN, LA, CT, MT.
Recall number
Z-0556-2021
Recall status
Open, Classified
CDRH record ID
184678
FDA record created
2020-12-10
Recall event number
86846

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗