Skip to content

Public FDA record

Z-0557-2019: STEPHANIX D2RS Digital Dynamic Remote Systemhttp://cts.fda.gov/division-tracking/images/trash.png Produc

Recall record Z-0557-2019 for Stephanix. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0557-2019
Company in FDA record
Stephanix
FDA record creation date
Dec 5, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has detected a potential risk using the command. After releasing the command, the movement of the table may continue instead of stopping.

Source facts

Action
The firm initiated the recall by email on 11/02/2018 to the sole US distributor. This email included a letter to be provided to the end user. The letter directed the user as follows: "As a precautionary measure and before your usual maintenance team intervention, the user should look at the equipment during all of its movement (as stated in the user manual). Immediately release the command if an abnormal movement is detected. If necessary, use the closest emergency switch to stop the system." f you need more information or technical assistance, you can phone your usual contact.
Initiated
2018-11-02
Root cause
Software design
Terminated
2021-07-25
Distribution
US Nationwide Distribution in the states of FL, IN, WV, TN, MO, WI, CA, FL, GA, PA
Recall number
Z-0557-2019
Recall status
Terminated
CDRH record ID
169013
FDA record created
2018-12-05
Recall event number
81582

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗