Skip to content

Public FDA record

Z-0557-2022: Chameleon PTA Balloon Dilation Catheter 8 mm x 40 mm, Chameleon PTA Balloon Catheter with Injection Port

Recall record Z-0557-2022 for Covidien Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0557-2022
Company in FDA record
Covidien Llc
FDA classification date
Feb 3, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to remove the protector sleeve). If the balloon is damaged while removing the protector sleeve and the catheter is subsequently used in a patient procedure, patient harm may potentially occur, including perforation of vessels, foreign body in patient, embolism, risks associated with surgical intervention, or delay to treatment/therapy.

Source facts

Action
On December 17, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers. Patient recommendation: Continue to monitor patients according to your hospital's patient management process. Required Actions: 1. Please immediately quarantine and discontinue use of the affected item code. 2. Please return affected product as indicated below. All products from the affected item code and lots listed below must be returned. 3. If you have distributed the products listed above, please promptly forward the information from this letter to those recipients. 4. Please complete the Recalled Product Return Form even if you do not have inventory. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Initiated
2021-12-17
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Germany, Israel, Italy, Portugal, Spain, Chile, Singapore, and China.
Recall number
Z-0557-2022
Recall status
Open, Classified
CDRH record ID
191509
FDA classified
2022-02-03
Recall event number
89418

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗