Open original FDA record ↗Recall record
- FDA identifier
- Z-0557-2022
- Company in FDA record
- Covidien Llc
- FDA classification date
- Feb 3, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to remove the protector sleeve). If the balloon is damaged while removing the protector sleeve and the catheter is subsequently used in a patient procedure, patient harm may potentially occur, including perforation of vessels, foreign body in patient, embolism, risks associated with surgical intervention, or delay to treatment/therapy.
Source facts
- Action
- On December 17, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers. Patient recommendation: Continue to monitor patients according to your hospital's patient management process. Required Actions: 1. Please immediately quarantine and discontinue use of the affected item code. 2. Please return affected product as indicated below. All products from the affected item code and lots listed below must be returned. 3. If you have distributed the products listed above, please promptly forward the information from this letter to those recipients. 4. Please complete the Recalled Product Return Form even if you do not have inventory. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
- Initiated
- 2021-12-17
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution - US Nationwide and the countries of Germany, Israel, Italy, Portugal, Spain, Chile, Singapore, and China.
- Recall number
- Z-0557-2022
- Recall status
- Open, Classified
- CDRH record ID
- 191509
- FDA classified
- 2022-02-03
- Recall event number
- 89418