Z-0557-2025: Tempus LS-Manual Defibrillator Model 00-3020 The Tempus LS-Manual is a defibrillator that can deliver a

Recall record Z-0557-2025 for Remote Diagnostic Technologies Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0557-2025
Company in FDA record
Remote Diagnostic Technologies Ltd.
FDA classification date
Nov 26, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Due to incorrect labels applied to devices during servicing. Labels contain the incorrect product name and CAT number.

Source facts

Action
On 10/28/2024, Philips sent an "URGENT Medical Device Correction" Letter via Certified Mail to customers informing them that SCHILLER AG, the manufacturer of the Tempus LS Manual Defibrillator, informed Philips that the incorrect labels were supplied with the service kit (kits 3, 6 and 8). The affected labels can be identified by the wrong device name (TEMPUS LS instead of TEMPUS LS MANUAL), the wrong CAT number (00-3010 instead of 00-3020) and the additional CE marking. SCHILLER AG initiated a correction on 10/07/2024. Customers are instructed to: Share this recall notice with those who need to be aware within their organization and any organization where the potentially affected devices have been transferred. It is recommended that the recall notice be posted on or near affected systems as a constant reminder of the issue until it is corrected. A Philips representative will contact affected customers to arrange for the corrected labels to be affixed to the affected Tempus LS Manual Defibrillators. For questions or assistance, contact Philips at 1-800-722-9377 or email ECR.Recall.Repsonse@Philips.com
Initiated
2024-10-28
Root cause
Labeling Change Control
Distribution
U.S. Nationwide distribution in the states of AL, CA, FL, MO, NY, OH, TX and VA.
Recall number
Z-0557-2025
Recall status
Open, Classified
CDRH record ID
210868
FDA classified
2024-11-26
Recall event number
95642

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗