Z-0558-2019: Artis Zee and Zeego, model no. 10272462 AXIOM Artis zee/Zeego is a family of dedicated angiography syste

Recall record Z-0558-2019 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0558-2019
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Dec 5, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This recall has been initiated due to a possible leakage of a hose in the cooling unit of the X-ray tube. The Artis zee system may lose coolant due to a possible leakage in a hose of the cooling unit in the X-ray tube. This will cause the X-ray tube not to be sufficiently cooled down. As a result, the system will display a message Tube hot, have a break . The X-ray will be blocked several minutes later and to prevent any damage to the machine the following message will be displayed NO XRAY, TUBE TOO HOT . This issue is very sporadic and may occur during an ongoing clinical procedure. Use of this product may have a potential impact on planned procedures as they may be delayed as well as in emergent cases, as clinical treatment may need to be terminated and transferred to a functioning system.

Source facts

Action
On October 26, 2018, the firm, Siemens Healthineers, distributed "URGENT: MEDICAL DEVICE CORRECTION" letters to affected customers. The letters informed customers of the product, problem and actions to be taken. The customers were instructed to do the following: promptly notify and instruct accordingly all the staff at your organization who need to be aware of this notice and Immediately examine your inventory and quarantine product subject to recall (if applicable). In addition, if you may have further distributed this product, please identify your customers and notify them at once of this product recall.If the device has been sold and is therefore no longer in your possession, please forward this notice to the new owner and inform us of the identity of the device's new owner where possible. The letter also stated that Siemens will replace the coolant pump on affected lots of cooling units to correct the issue. Siemens' service organization will contact customers to arrange a date to perform the corrective action with Update Instruction AX013/18/S. To request an earlier appointment, feel free to contact the service organization at 1-800-888-7436. If you have any questions, contact Regulatory Technical Specialist at 610-219-4834 or email to: anastasia.sokolova@siemens-healthineers.com.
Initiated
2018-10-26
Root cause
Process control
Terminated
2020-05-12
Distribution
US Distribution to states of: AR, CA, KS, KY, MD, MN, NM, OK, PA, RI, TX, and UT.
Recall number
Z-0558-2019
Recall status
Terminated
CDRH record ID
168937
FDA record created
2018-12-05
Recall event number
81545

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗