Skip to content

Public FDA record

Z-0558-2021: MANI Infusion Cannula 23G S - Product Usage: for use in injecting perfusate or aspirating out intraocular

Recall record Z-0558-2021 for Mani, Inc. - Kiyohara Facility. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0558-2021
Company in FDA record
Mani, Inc. - Kiyohara Facility
FDA record creation date
Dec 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Dimensions of the Luer connector parts out of specification, may cause leakage of fluid and unintended disconnection during use

Source facts

Action
MANI Inc. issued Voluntary Removal letter issued on 10/27/20 via email to the US Distributor stating reason for removal, health risk and action to take: request distributors to return the devices. If there are MANI TROCAR KIT and MANI INFUSION CANNULA in your stock and/or your customer's stock: Please write the returning Q'ty. in the list ,"Returned" If MANI TROCAR KIT and MANI INFUSION CANNULA are already used up : Please write the used Q'ty. in the list , "Used. Please send this sheet and relevant items if you have to our address below. Any costs/fees related to this return shipment should be charged to MANI FedEx account. MANI, INC. 8-3, Kiyohara Industrial Park, Utsunomiya, Tochigi, 321-3231, Japan
Initiated
2020-10-27
Root cause
Unknown/Undetermined by firm
Terminated
2021-08-11
Distribution
US Nationwide distribution in the state of MO.
Recall number
Z-0558-2021
Recall status
Terminated
CDRH record ID
184735
FDA record created
2020-12-10
Recall event number
86873

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗