- Action
- Philips issued Urgent Medical Device Correction (UMDC) letter US customers in the US on 11/3/23 and a Field Safety Notice (FSN) to customers outside the US. Letter states reason for recall, health risk and action to take:
As a remedy, Philips is replacing the existing covers of all the affected Philips Allura and Philips Azurion
with the monoplane fixed ceiling mounted systems with a redesigned L-arm rotation cover (bolted
instead of spring clips) to make sure it is resistant to external collisions.
Philips will contact all affected customers to arrange for a Field Service Engineer visit to replace the Larm
rotation cover free of charge (reference FCO72200510 for the Philips Allura systems including the
Philips Poly G H5000, FCO72200511 for the Philips Azurion systems).
If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
- Initiated
- 2023-11-03
- Root cause
- Device Design
- Distribution
- Worldwide - US Nationwide and the countries of Albania, Algeria, Argentina, Australia,
Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, CANADA, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic,
Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Gabon, Georgia,
Germany, Ghana, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong,
Hungary, Iceland, India, Indonesia, Iran, IRAQ, Ireland, Israel, Italy, Japan, Jordan,
Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Latvia, Latvia, Republic of
Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Martinique, Mauritius,
Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand,
Nicaragua, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea,
Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania,
Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan,
Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay,
Uzbekistan, Venezuela, Vietnam, Yemen.
- Recall number
- Z-0558-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204295
- FDA classified
- 2023-12-15
- Recall event number
- 93416