Skip to content

Public FDA record

Z-0559-2023: NUVASIVE MAGEC MAGEC 2 Rod. The MAGEC System is used to brace the spine during growth to minimize the p

Recall record Z-0559-2023 for NuVasive Specialized Orthopedics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0559-2023
Company in FDA record
NuVasive Specialized Orthopedics, Inc.
FDA classification date
Dec 14, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During accelerated shelf-life extension studies it was noted that the final packaging was experiencing a potential of failure of the sterile packaging barrier which could result in a hazardous situation where the device is no longer sterile.

Source facts

Action
On 11/01/2022, the firm sent an URGENT VOLUNTARY MEDICAL DEVICE RECALL Letter via Docusign/FedEx/email to customers (US/OUS Surgeons and Distributors) informing them that the packaging closure integrity (manufactured with an undersized O-ring of the MAGEC 2 (marketed as MAGEC X) packaged units. As a result of the undersized O-ring gland diameter, the O-ring compression is affected when the hard cap if affixed to the packaging tube. The lack of compression can create a situation where the integrity of the container closure system is compromised potentially impacting the sterile barrier system. Customer are instructed to: 1. Immediately quarantine any affected product. 2. Return all affected products. The NuVasive sales representative will be contacting customers to assist with the return actions. 3. Review this notification to ensure that affected personnel are aware of its contents. 4. For any devices that have been implanted, there are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedules. For questions or assistance - contact Sr. Director, Global Quality at 858-344-2667, email pyrigoyen@nuvasive.com, or the NuVasive Sales Representative.
Initiated
2022-11-01
Root cause
Process control
Distribution
Distribution US nationwide including Puerto Rico; Chile, Israel, Jordan, New Zealand, Singapore, Sweden, Thailand, and United Arab Emirates.
Recall number
Z-0559-2023
Recall status
Open, Classified
CDRH record ID
196860
FDA classified
2022-12-14
Recall event number
91125

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗